Standard Features:
With 5.5" SNT screen. Mounted on 4 castors. 2 front castors with brakes.
With 2 big drawers.
Handle for easy maneuvering.
Epoxy painted high quality mild steel frame.
Equipped with High precise Anesthesia Vaporizer (with temperature, flow and pressure compensation function).
The agent output is continuous, stable & precise. The agent type: Isoflurane, Enflurane, Halothane, Sevoflurane
Anesthesia ventilator with IPPV and SIMV mode. The ventilator can display P-t and V-t wave-form at the same time.
Necessary monitoring and alarm function to ensure operational safety.
Classical CO2 circle absorber with 2 canisters for low resistance and minimum gas leakage.
Optional : O2 , N2 O Yoke device. When main gas supply fails, the backup gas supply starts.
Feature :
Patient type :
Adult, Pediatric
Driven mode :
Pneumatically driven electronically controlled
Mode :
Semi-open, Semi-closed or Closed system
Anesthetic vaporizer :
One vaporizer
Power supply :
220V+/-10%, 50Hz+/-1Hz
Power consumption :
Not more than 50W
Internal battery :
DC12V (over 4 working hours)
Anesthesia Ventilator Specification :
Respiration mode :
IPPV, SIMV, SIGH, MANUAL
Waveform :
P-t, V-t
Tidal volume :
50ml~1600ml
Respiratory frequency :
1~65 BPM / I: E ratio: 8: 1~1: 8
Alarm parameters :
High pressure alarm, Low pressure alarm, O2 source failure alarm
High pressure alarm limit:
20~60CMH2 O / Low pressure alarm limit: 0~20CMH2 O
Monitor parameters :
Tidal volume, MV, Ppeak, PEEP, respiratory frequency, I: E ratio
Safety valve pressure limit :
6 Kpa
Main Unit :
Circle absorber :
Compact Circle absorber Bellow : 0 ~ 1600ml
Flow meter :
O2 , N2 O four-tube flow meter
O2 tube: 0L/min~1L/min~10L/min
N2 O tube: 0L/min~1L/min~10L/min
O2 and N2 O linkage device
Adjust O2 and N2 O in proportion ( keep O2 concentration not lower than 25% )
Oxygen Flush :
25L/min ~ 75L/min
Gas source pressure :
O2 : 0.25 ~ 0.65Mpa
N2 O: 0.25~ 0.65Mpa
Gauge :
2 units, one N2 O, one O2