Infant Radiant Warmer, Open Care System - Apple Series



NIC201a - Infant Radiant Warmer with Fixed Baby Cradle and Drawers with Microprocessor based temperature controller with 4 modes Skin / Air / Manual / Prewarm mode
Standard Specifications for NIC201a
  • Fixed height warmer with attached baby cradle and drawers mounted on 4" (100mm) wheels, 2 with brake.
  • Stainless steel baby tray with water proof, zipped cover mattress of density approx. 21-25 kg/m³
  • Continuous head up/down facility
  • X-Ray cassette holder
  • Collapsible/lockable Side Supports made up of Unbreakable polycarbonate material
  • Slots for sensor and ventilator tubing
  • Side tray for placement of monitor
  • Stainless steel infusion rod with facility to mount Syringe/infusion pump
  • Lower shelf below baby tray to facilitate under surface phototherapy, Facility of 3 drawers for storage
  • Height : 74.5" (1890 mm), Width : 28.0" (710 mm), Length : 39.0" (990 mm)
  • Bed to Floor Height : 37.5" (950 mm)
  • Distance between baby bed and heater: 33" (840 mm)
  • Mattress size: 30" (762 mm) x 20" (508 mm)
  • Height of side panel above baby bed: 5.75" (145 mm)
Performance/Electrical Specifications :
  • Microprocessor based temperature controller with Pre-warm, Manual and Skin Servo modes
  • LED Display of Skin Temp.,Air Temp., Set Temp. (Resolution 0.1°C) & Heater Power %.
  • LCD to display modes of operation, Alarm messages, Skin Temperature in °F and Timer.
  • Thermistor based sensor with Accuracy of +/- 0.2°C (Temp. Calibration not require)
  • Modes of Operation: Skin Servo, Air, Manual and Pre-Warm
  • Keypad lock facility
  • Skin mode: Temperature range: 34 to 38°C , Resolution : 0.1°C
  • Air mode: Temperature range: Room temp. to 39°C , Resolution : 0.1°C
  • Manual mode: Adjustable Heater Power 0 to 100% (with an increment of 5%), Time duration:15mins
  • Pre-warm mode: To pre-warm the cradle before baby is placed on the bed. (No alarm will be activated during Prewarm mode) Adjustable Heater Power 0 to 25% (with an increment of 5%), Time duration:30mins
  • Overhead heater box: Swivelling, Consists of Ceramic heater (650 Watts) placed in parabolic reflector with a safety grill & Halogen observation light (12V/50W) for observing the baby
  • Operating Voltage: 220-240 VAC, 50Hz, Power: 800W (Max.)
  • Alarm: Audio alarm with 10 minute mute facility & Visual alarms with message on LCD as follows
Alarm Condition Alarm Messages on LCD
Skin Sensor is unplugged /Faulty Skin Sensor Fail
Air Sensor is unplugged /Faulty Air Sensor Fail
Measured skin temperature higher than set temperature by 1.0°C Skin Temp. High
Measured skin temperature lower than set temperature by 1.0°C Skin Temp. Low
Measured skin temperature higher than 38.0°C (With cut-off) Skin Over Range
Measured skin temperature lower than 34.0°C Skin Under Range
Measured air temperature higher than set temperature by 1.5°C Air Temp. High
Measured air temperature lower than set temperature by 3.0°C Air Temp. Low
Measured air temperature higher than 39.0°C (With cut-off) Air Over Range
Heater power is 100% for more than 15mins Chk Sens Position
Heater is unplugged or faulty Heater Open
Heater Power drive fail/uncontrollable Chk. Pwr. Drive
Power Fail Alarm: Audio with LED indicator

NIC211a - Infant Radiant Warmer with Fixed Baby Cradle and Drawers with Microprocessor based temperature controller with 3 modes Skin / Manual / Prewarm mode
Standard Specifications for NIC211a
  • Fixed height warmer with attached baby cradle and drawers mounted on 4" (100mm) wheels, 2 with brake.
  • Stainless steel baby tray with water proof, zipped cover mattress of density approx. 21-25 kg/m³
  • Continuous head up/down facility
  • X-Ray cassette holder
  • Collapsible/lockable Side Supports made up of Unbreakable polycarbonate material
  • Slots for sensor and ventilator tubing
  • Side tray for placement of monitor
  • Stainless steel infusion rod with facility to mount Syringe/infusion pump
  • Lower shelf below baby tray to facilitate under surface phototherapy, Facility of 3 drawers for storage
  • Height : 74.5" (1890 mm), Width : 28.0" (710 mm), Length : 39.0" (990 mm)
  • Bed to Floor Height : 37.5" (950 mm)
  • Distance between baby bed and heater: 33" (840 mm)
  • Mattress size: 30" (762 mm) x 20" (508 mm)
  • Height of side panel above baby bed: 5.75" (145 mm)
Performance/Electrical Specifications :
  • Microprocessor based temperature controller with Pre-warm, Manual and Skin Servo modes
  • LED Display of Skin Temp., Set Temp. (Resolution 0.1°C) & Heater Power %.
  • LCD to display modes of operation, Alarm messages, Skin Temperature in °F and Timer.
  • Thermistor based sensor with Accuracy of +/- 0.2°C (Temp. Calibration not require)
  • Modes of Operation: Skin Servo, Manual and Pre-Warm
  • Keypad lock facility
  • Skin mode: Temperature range: 34 to 38°C , Resolution : 0.1°C
  • Manual mode: Adjustable Heater Power 0 to 100% (with an increment of 5%), Time duration:15mins
  • Pre-warm mode: To pre-warm the cradle before baby is placed on the bed. (No alarm will be activated during Prewarm mode). Adjustable Heater Power 0 to 25% (with an increment of 5%), Time duration:30mins
  • Overhead heater box: Swivelling, Consists of Ceramic heater (650 Watts) placed in parabolic reflector with a safety grill & Halogen observation light (12V/50W) for observing the baby
  • Operating Voltage: 220-240 VAC, 50Hz, Power: 800W (Max.)
  • Alarm: Audio alarm with 10 minute mute facility & Visual alarms with message on LCD as follows
Alarm Condition Alarm Messages on LCD
Skin Sensor is unplugged /Faulty Skin Sensor Fail
Measured skin temperature higher than set temperature by 1.0°C Skin Temp. High
Measured skin temperature lower than set temperature by 1.0°C Skin Temp. Low
Measured skin temperature higher than 38.0°C (With cut-off) Skin Over Range
Measured skin temperature lower than 34.0°C Skin Under Range
Heater power is 100% for more than 15mins Chk Sens Position
Heater is unplugged or faulty Heater Open
Heater Power drive fail/uncontrollable Chk. Pwr. Drive
Power Fail Alarm: Audio with LED indicator

NIC202a - Infant Radiant Warmer with Fixed Baby Cradle and Drawers with Microprocessor based temperature controller with 4 modes Skin / Air / Manual / Prewarm mode
Standard Specifications for NIC202a
  • Fixed height warmer with attached baby cradle and drawers mounted on 4" (100mm) wheels, 2 with brake.
  • Stainless steel baby tray with water proof, zipped cover mattress of density approx. 21-25 kg/m³
  • Continuous head up/down facility
  • X-Ray cassette holder
  • Collapsible/lockable Side Supports made up of Unbreakable polycarbonate material
  • Slots for sensor and ventilator tubing
  • Side tray for placement of monitor
  • Stainless steel infusion rod with facility to mount Syringe/infusion pump
  • Lower shelf below baby tray to facilitate under surface phototherapy.
  • Height : 74.5" (1890 mm), Width : 28.0" (710 mm), Length : 39.0" (990 mm)
  • Bed to Floor Height : 37.5" (950 mm)
  • Distance between baby bed and heater: 33" (840 mm)
  • Mattress size: 30" (762 mm) x 20" (508 mm)
  • Height of side panel above baby bed: 5.75" (145 mm)
Performance/Electrical Specifications :
  • Microprocessor based temperature controller with Pre-warm, Manual and Skin Servo modes
  • LED Display of Skin Temp.,Air Temp., Set Temp. (Resolution 0.1°C) & Heater Power %.
  • LCD to display modes of operation, Alarm messages, Skin Temperature in °F and Timer.
  • Thermistor based sensor with Accuracy of +/- 0.2°C (Temp. Calibration not require)
  • Modes of Operation: Skin Servo, Air, Manual and Pre-Warm
  • Keypad lock facility
  • Skin mode: Temperature range: 34 to 38°C , Resolution : 0.1°C
  • Air mode: Temperature range: Room temp. to 39°C , Resolution : 0.1°C
  • Manual mode: Adjustable Heater Power 0 to 100% (with an increment of 5%), Time duration:15mins
  • Pre-warm mode: To pre-warm the cradle before baby is placed on the bed. (No alarm will be activated during Prewarm mode) Adjustable Heater Power 0 to 25% (with an increment of 5%), Time duration:30mins
  • Overhead heater box: Swivelling, Consists of Ceramic heater (650 Watts) placed in parabolic reflector with a safety grill & Halogen observation light (12V/50W) for observing the baby
  • Operating Voltage: 220-240 VAC, 50Hz, Power: 800W (Max.)
  • Alarm: Audio alarm with 10 minute mute facility & Visual alarms with message on LCD as follows
Alarm Condition Alarm Messages on LCD
Skin Sensor is unplugged /Faulty Skin Sensor Fail
Air Sensor is unplugged /Faulty Air Sensor Fail
Measured skin temperature higher than set temperature by 1.0°C Skin Temp. High
Measured skin temperature lower than set temperature by 1.0°C Skin Temp. Low
Measured skin temperature higher than 38.0°C (With cut-off) Skin Over Range
Measured skin temperature lower than 34.0°C Skin Under Range
Measured air temperature higher than set temperature by 1.5°C Air Temp. High
Measured air temperature lower than set temperature by 3.0°C Air Temp. Low
Measured air temperature higher than 39.0°C (With cut-off) Air Over Range
Heater power is 100% for more than 15mins Chk Sens Position
Heater is unplugged or faulty Heater Open
Heater Power drive fail/uncontrollable Chk. Pwr. Drive
Power Fail Alarm: Audio with LED indicator

NIC212a - Infant Radiant Warmer with Fixed Baby Cradle and Drawers with Microprocessor based temperature controller with 3 modes Skin / Manual / Prewarm mode
Standard Specifications for NIC212a
  • Fixed height warmer with attached baby cradle mounted on 4" (100mm) wheels, 2 with brake.
  • Stainless steel baby tray with water proof, zipped cover mattress of density approx. 21-25 kg/m³
  • Continuous head up/down facility
  • X-Ray cassette holder
  • Collapsible/lockable Side Supports made up of Unbreakable polycarbonate material
  • Slots for sensor and ventilator tubing
  • Side tray for placement of monitor
  • Stainless steel infusion rod with facility to mount Syringe/infusion pump
  • Lower shelf below baby tray to facilitate under surface phototherapy.
  • Height : 74.5" (1890 mm), Width : 28.0" (710 mm), Length : 39.0" (990 mm)
  • Bed to Floor Height : 37.5" (950 mm)
  • Distance between baby bed and heater: 33" (840 mm)
  • Mattress size: 30" (762 mm) x 20" (508 mm)
  • Height of side panel above baby bed: 5.75" (145 mm)
Performance/Electrical Specifications :
  • Microprocessor based temperature controller with Pre-warm, Manual and Skin Servo modes
  • LED Display of Skin Temp., Set Temp. (Resolution 0.1°C) & Heater Power %.
  • LCD to display modes of operation, Alarm messages, Skin Temperature in °F and Timer.
  • Thermistor based sensor with Accuracy of +/- 0.2°C (Temp. Calibration not require)
  • Modes of Operation: Skin Servo, Manual and Pre-Warm
  • Keypad lock facility
  • Skin mode: Temperature range: 34 to 38°C , Resolution : 0.1°C
  • Manual mode: Adjustable Heater Power 0 to 100% (with an increment of 5%), Time duration:15mins
  • Pre-warm mode: To pre-warm the cradle before baby is placed on the bed. (No alarm will be activated during Prewarm mode). Adjustable Heater Power 0 to 25% (with an increment of 5%), Time duration:30mins
  • Overhead heater box: Swivelling, Consists of Ceramic heater (650 Watts) placed in parabolic reflector with a safety grill & Halogen observation light (12V/50W) for observing the baby
  • Operating Voltage: 220-240 VAC, 50Hz, Power: 800W (Max.)
  • Alarm: Audio alarm with 10 minute mute facility & Visual alarms with message on LCD as follows
Alarm Condition Alarm Messages on LCD
Skin Sensor is unplugged /Faulty Skin Sensor Fail
Measured skin temperature higher than set temperature by 1.0°C Skin Temp. High
Measured skin temperature lower than set temperature by 1.0°C Skin Temp. Low
Measured skin temperature higher than 38.0°C (With cut-off) Skin Over Range
Measured skin temperature lower than 34.0°C Skin Under Range
Heater power is 100% for more than 15mins Chk Sens Position
Heater is unplugged or faulty Heater Open
Heater Power drive fail/uncontrollable Chk. Pwr. Drive
Power Fail Alarm: Audio with LED indicator

NIC203a - Infant Radiant Warmer with Detachable Baby Cradle with Microprocessor based temperature controller with 4 modes Skin / Air / Manual / Prewarm mode
Standard Specifications
  • Fixed height warmer with detachable baby cradle mounted on 4" (100mm) wheels, 2 with brake.
  • Stainless steel baby tray with water proof, zipped cover mattress of density approx. 21-25 kg/m³
  • Continuous head up/down facility
  • X-Ray cassette holder
  • Collapsible/lockable Side Supports made up of Unbreakable polycarbonate material
  • Slots for sensor and ventilator tubing
  • Side tray for placement of monitor
  • Stainless steel infusion rod with facility to mount Syringe/infusion pump
  • Lower shelf below baby tray to facilitate under surface phototherapy.
  • Height : 74.5" (1890 mm), Width : 28.0" (710 mm), Length : 39.0" (990 mm)
  • Bed to Floor Height : 37.5" (950 mm)
  • Distance between baby bed and heater: 33" (840 mm)
  • Mattress size: 30" (762 mm) x 20" (508 mm)
  • Height of side panel above baby bed: 5.75" (145 mm)
Performance/Electrical Specifications :
  • Microprocessor based temperature controller with Pre-warm, Manual and Skin Servo modes
  • LED Display of Skin Temp., Air Temp., Set Temp. (Resolution 0.1°C) & Heater Power %.
  • LCD to display modes of operation, Alarm messages, Skin Temperature in °F and Timer.
  • Thermistor based sensor with Accuracy of +/- 0.2°C (Temp. Calibration not require)
  • Modes of Operation: Skin Servo, Manual and Pre-Warm
  • Keypad lock facility
  • Skin mode: Temperature range: 34 to 38°C , Resolution : 0.1°C
  • Manual mode: Adjustable Heater Power 0 to 100% (with an increment of 5%), Time duration:15mins
  • Pre-warm mode: To pre-warm the cradle before baby is placed on the bed. (No alarm will be activated during Prewarm mode). Adjustable Heater Power 0 to 25% (with an increment of 5%), Time duration:30mins
  • Overhead heater box: Swivelling, Consists of Ceramic heater (650 Watts) placed in parabolic reflector with a safety grill & Halogen observation light (12V/50W) for observing the baby
  • Operating Voltage: 220-240 VAC, 50Hz, Power: 800W (Max.)
  • Alarm: Audio alarm with 10 minute mute facility & Visual alarms with message on LCD as follows
Alarm Condition Alarm Messages on LCD
Skin Sensor is unplugged /Faulty Skin Sensor Fail
Air Sensor is unplugged /Faulty Air Sensor Fail
Measured skin temperature higher than set temperature by 1.0°C Skin Temp. High
Measured skin temperature lower than set temperature by 1.0°C Skin Temp. Low
Measured skin temperature higher than 38.0°C (With cut-off) Skin Over Range
Measured skin temperature lower than 34.0°C Skin Under Range
Measured air temperature higher than set temperature by 1.5°C Air Temp. High
Measured air temperature lower than set temperature by 3.0°C Air Temp. Low
Measured air temperature higher than 39.0°C (With cut-off) Air Over Range
Heater power is 100% for more than 15mins Chk Sens Position
Heater is unplugged or faulty Heater Open
Heater Power drive fail/uncontrollable Chk. Pwr. Drive
Power Fail Alarm: Audio with LED indicator

NIC213a - Infant Radiant Warmer with Detachable Baby Cradle and Drawers with Microprocessor based temperature controller with 3 modes Skin / Manual / Prewarm mode
Standard Specifications
  • Fixed height warmer with detachable baby cradle mounted on 4" (100mm) wheels, 2 with brake.
  • Stainless steel baby tray with water proof, zipped cover mattress of density approx. 21-25 kg/m³
  • Continuous head up/down facility
  • X-Ray cassette holder
  • Collapsible/lockable Side Supports made up of Unbreakable polycarbonate material
  • Slots for sensor and ventilator tubing
  • Side tray for placement of monitor
  • Stainless steel infusion rod with facility to mount Syringe/infusion pump
  • Lower shelf below baby tray to facilitate under surface phototherapy.
  • Height : 74.5" (1890 mm), Width : 28.0" (710 mm), Length : 39.0" (990 mm)
  • Bed to Floor Height : 37.5" (950 mm)
  • Distance between baby bed and heater: 33" (840 mm)
  • Mattress size: 30" (762 mm) x 20" (508 mm)
  • Height of side panel above baby bed: 5.75" (145 mm)
Performance/Electrical Specifications :
  • Microprocessor based temperature controller with Pre-warm, Manual and Skin Servo modes
  • LED Display of Skin Temp., Set Temp. (Resolution 0.1°C) & Heater Power %.
  • LCD to display modes of operation, Alarm messages, Skin Temperature in °F and Timer.
  • Thermistor based sensor with Accuracy of +/- 0.2°C (Temp. Calibration not require)
  • Modes of Operation: Skin Servo, Manual and Pre-Warm
  • Keypad lock facility
  • Skin mode: Temperature range: 34 to 38°C , Resolution : 0.1°C
  • Manual mode: Adjustable Heater Power 0 to 100% (with an increment of 5%), Time duration:15mins
  • Pre-warm mode: To pre-warm the cradle before baby is placed on the bed. (No alarm will be activated during Prewarm mode). Adjustable Heater Power 0 to 25% (with an increment of 5%), Time duration:30mins
  • Overhead heater box: Swivelling, Consists of Ceramic heater (650 Watts) placed in parabolic reflector with a safety grill & Halogen observation light (12V/50W) for observing the baby
  • Operating Voltage: 220-240 VAC, 50Hz, Power: 800W (Max.)
  • Alarm: Audio alarm with 10 minute mute facility & Visual alarms with message on LCD as follows
Alarm Condition Alarm Messages on LCD
Skin Sensor is unplugged /Faulty Skin Sensor Fail
Measured skin temperature higher than set temperature by 1.0°C Skin Temp. High
Measured skin temperature lower than set temperature by 1.0°C Skin Temp. Low
Measured skin temperature higher than 38.0°C (With cut-off) Skin Over Range
Measured skin temperature lower than 34.0°C Skin Under Range
Heater power is 100% for more than 15mins Chk Sens Position
Heater is unplugged or faulty Heater Open
Heater Power drive fail/uncontrollable Chk. Pwr. Drive
Power Fail Alarm: Audio with LED indicator

NIC205 - Infant Radiant Warmer with Fixed Baby Cradle with Microprocessor based temperature controller with 4 modes Skin / Air / Manual / Prewarm mode
Standard Specifications
  • Fixed height warmer with attachable baby cradle and drawers mounted on 4" (100mm) wheels, 2 with brake.
  • Stainless steel baby tray with water proof, zipped cover mattress of density approx. 21-25 kg/m³
  • Continuous head up/down facility
  • X-Ray cassette holder
  • Collapsible/lockable Side Supports made up of Unbreakable polycarbonate material
  • Slots for sensor and ventilator tubing
  • Side tray for placement of monitor
  • Stainless steel infusion rod with facility to mount Syringe/infusion pump
  • Lower shelf below baby tray to facilitate under surface phototherapy, Facility of 3 drawers for storage
  • Height : 74.5" (1890 mm), Width : 28.0" (710 mm), Length : 39.0" (990 mm)
  • Bed to Floor Height : 37.5" (950 mm)
  • Distance between baby bed and heater: 33" (840 mm)
  • Mattress size: 30" (762 mm) x 20" (508 mm)
  • Height of side panel above baby bed: 5.75" (145 mm)
Performance/Electrical Specifications :
  • Microprocessor based temperature controller with Pre-warm, Manual and Skin Servo modes
  • LED Display of Skin Temp., Air Temp., Set Temp. (Resolution 0.1°C) & Heater Power %.
  • LCD to display modes of operation, Alarm messages, Skin Temperature in °F and Timer.
  • Thermistor based sensor with Accuracy of +/- 0.2°C (Temp. Calibration not require)
  • Modes of Operation: Skin Servo, Air, Manual and Pre-Warm
  • Keypad lock facility
  • Skin mode: Temperature range: 34 to 38°C , Resolution : 0.1°C
  • Air mode: Temperature range: Room temp. to 39°C , Resolution : 0.1°C
  • Manual mode: Adjustable Heater Power 0 to 100% (with an increment of 5%), Time duration:15mins
  • Pre-warm mode: To pre-warm the cradle before baby is placed on the bed. (No alarm will be activated during Prewarm mode). Adjustable Heater Power 0 to 25% (with an increment of 5%), Time duration:30mins
  • Overhead heater box: Swivelling, Consists of Ceramic heater (650 Watts) placed in parabolic reflector with a safety grill & Halogen observation light (12V/50W) for observing the baby
  • Operating Voltage: 220-240 VAC, 50Hz, Power: 800W (Max.)
  • Alarm: Audio alarm with 10 minute mute facility & Visual alarms with message on LCD as follows
Alarm Condition Alarm Messages on LCD
Skin Sensor is unplugged /Faulty Skin Sensor Fail
Air Sensor is unplugged /Faulty Air Sensor Fail
Measured skin temperature higher than set temperature by 1.0°C Skin Temp. High
Measured skin temperature lower than set temperature by 1.0°C Skin Temp. Low
Measured skin temperature higher than 38.0°C (With cut-off) Skin Over Range
Measured skin temperature lower than 34.0°C Skin Under Range
Measured air temperature higher than set temperature by 1.5°C Air Temp. High
Measured air temperature lower than set temperature by 3.0°C Air Temp. Low
Measured air temperature higher than 39.0°C (With cut-off) Air Over Range
Heater power is 100% for more than 15mins Chk Sens Position
Heater is unplugged or faulty Heater Open
Heater Power drive fail/uncontrollable Chk. Pwr. Drive
Power Fail Alarm: Audio with LED indicator

NIC207 - Infant Radiant Warmer with Fixed Baby Cradle with Microprocessor based temperature controller with 4 modes Skin / Air / Manual / Prewarm mode
Standard Specifications
  • Fixed height warmer with attached baby cradle mounted on 4" (100mm) wheels, 2 with brake.
  • Stainless steel baby tray with water proof, zipped cover mattress of density approx. 21-25 kg/m³
  • Continuous head up/down facility
  • X-Ray cassette holder
  • Collapsible/lockable Side Supports made up of Unbreakable polycarbonate material
  • Slots for sensor and ventilator tubing
  • Side tray for placement of monitor
  • Stainless steel infusion rod with facility to mount Syringe/infusion pump
  • Lower shelf below baby tray to facilitate under surface phototherapy.
  • Height : 74.5" (1890 mm), Width : 28.0" (710 mm), Length : 39.0" (990 mm)
  • Bed to Floor Height : 37.5" (950 mm)
  • Distance between baby bed and heater: 33" (840 mm)
  • Mattress size: 30" (762 mm) x 20" (508 mm)
  • Height of side panel above baby bed: 5.75" (145 mm)
Performance/Electrical Specifications :
  • Microprocessor based temperature controller with Pre-warm, Manual and Skin Servo modes
  • LED Display of Skin Temp., Air Temp., Set Temp. (Resolution 0.1°C) & Heater Power %.
  • LCD to display modes of operation, Alarm messages, Skin Temperature in °F and Timer.
  • Thermistor based sensor with Accuracy of +/- 0.2°C (Temp. Calibration not require)
  • Modes of Operation: Skin Servo, Air, Manual and Pre-Warm
  • Keypad lock facility
  • Skin mode: Temperature range: 34 to 38°C , Resolution : 0.1°C
  • Air mode: Temperature range: Room temp. to 39°C , Resolution : 0.1°C
  • Manual mode: Adjustable Heater Power 0 to 100% (with an increment of 5%), Time duration:15mins
  • Pre-warm mode: To pre-warm the cradle before baby is placed on the bed. (No alarm will be activated during Prewarm mode). Adjustable Heater Power 0 to 25% (with an increment of 5%), Time duration:30mins
  • Overhead heater box: Swivelling, Consists of Ceramic heater (650 Watts) placed in parabolic reflector with a safety grill & Halogen observation light (12V/50W) for observing the baby
  • Operating Voltage: 220-240 VAC, 50Hz, Power: 800W (Max.)
  • Alarm: Audio alarm with 10 minute mute facility & Visual alarms with message on LCD as follows
Alarm Condition Alarm Messages on LCD
Skin Sensor is unplugged /Faulty Skin Sensor Fail
Air Sensor is unplugged /Faulty Air Sensor Fail
Measured skin temperature higher than set temperature by 1.0°C Skin Temp. High
Measured skin temperature lower than set temperature by 1.0°C Skin Temp. Low
Measured skin temperature higher than 38.0°C (With cut-off) Skin Over Range
Measured skin temperature lower than 34.0°C Skin Under Range
Measured air temperature higher than set temperature by 1.5°C Air Temp. High
Measured air temperature lower than set temperature by 3.0°C Air Temp. Low
Measured air temperature higher than 39.0°C (With cut-off) Air Over Range
Heater power is 100% for more than 15mins Chk Sens Position
Heater is unplugged or faulty Heater Open
Heater Power drive fail/uncontrollable Chk. Pwr. Drive
Power Fail Alarm: Audio with LED indicator

NIC206 - Infant Radiant Warmer with Fixed Baby Cradle and Drawers with Microprocessor based temperature controller with 3 modes Skin / Manual / Prewarm mode
Standard Specifications
  • Fixed height warmer with attachable baby cradle and drawers mounted on 4" (100mm) wheels, 2 with brake.
  • Stainless steel baby tray with water proof, zipped cover mattress of density approx. 21-25 kg/m³
  • Continuous head up/down facility
  • X-Ray cassette holder
  • Collapsible/lockable Side Supports made up of Unbreakable polycarbonate material
  • Slots for sensor and ventilator tubing
  • Side tray for placement of monitor
  • Stainless steel infusion rod with facility to mount Syringe/infusion pump
  • Lower shelf below baby tray to facilitate under surface phototherapy, Facility of 3 drawers for storage
  • Height : 74.5" (1890 mm), Width : 28.0" (710 mm), Length : 39.0" (990 mm)
  • Bed to Floor Height : 37.5" (950 mm)
  • Distance between baby bed and heater: 33" (840 mm)
  • Mattress size: 30" (762 mm) x 20" (508 mm)
  • Height of side panel above baby bed: 5.75" (145 mm)
Performance/Electrical Specifications :
  • Microprocessor based temperature controller with Pre-warm, Manual and Skin Servo modes
  • LED Display of Skin Temp., Set Temp. (Resolution 0.1°C) & Heater Power %.
  • LCD to display modes of operation, Alarm messages, Skin Temperature in °F and Timer.
  • Thermistor based sensor with Accuracy of +/- 0.2°C (Temp. Calibration not require)
  • Modes of Operation: Skin Servo, Manual and Pre-Warm
  • Keypad lock facility
  • Skin mode: Temperature range: 34 to 38°C , Resolution : 0.1°C
  • Manual mode: Adjustable Heater Power 0 to 100% (with an increment of 5%), Time duration:15mins
  • Pre-warm mode: To pre-warm the cradle before baby is placed on the bed. (No alarm will be activated during Prewarm mode). Adjustable Heater Power 0 to 25% (with an increment of 5%), Time duration:30mins
  • Overhead heater box: Swivelling, Consists of Ceramic heater (650 Watts) placed in parabolic reflector with a safety grill & Halogen observation light (12V/50W) for observing the baby
  • Operating Voltage: 220-240 VAC, 50Hz, Power: 800W (Max.)
  • Alarm: Audio alarm with 10 minute mute facility & Visual alarms with message on LCD as follows
Alarm Condition Alarm Messages on LCD
Skin Sensor is unplugged /Faulty Skin Sensor Fail
Measured skin temperature higher than set temperature by 1.0°C Skin Temp. High
Measured skin temperature lower than set temperature by 1.0°C Skin Temp. Low
Measured skin temperature higher than 38.0°C (With cut-off) Skin Over Range
Measured skin temperature lower than 34.0°C Skin Under Range
Heater power is 100% for more than 15mins Chk Sens Position
Heater is unplugged or faulty Heater Open
Heater Power drive fail/uncontrollable Chk. Pwr. Drive
Power Fail Alarm: Audio with LED indicator

NIC208 - Infant Radiant Warmer with Fixed Baby Cradle with Microprocessor based temperature controller with 3 modes Skin / Manual / Prewarm mode
Standard Specifications
  • Fixed height with attached baby cradle mounted on 4" (100mm) wheels, 2 with brake.
  • Stainless steel baby tray with water proof, zipped cover mattress of density approx. 21-25 kg/m³
  • Continuous head up/down facility
  • X-Ray cassette holder
  • Collapsible/lockable Side Supports made up of Unbreakable polycarbonate material
  • Slots for sensor and ventilator tubing
  • Side tray for placement of monitor
  • Stainless steel infusion rod with facility to mount Syringe/infusion pump
  • Lower shelf below baby tray to facilitate under surface phototherapy.
  • Height : 74.5" (1890 mm), Width : 28.0" (710 mm), Length : 39.0" (990 mm)
  • Bed to Floor Height : 37.5" (950 mm)
  • Distance between baby bed and heater: 33" (840 mm)
  • Mattress size: 30" (762 mm) x 20" (508 mm)
  • Height of side panel above baby bed: 5.75" (145 mm)
Performance/Electrical Specifications :
  • Microprocessor based temperature controller with Pre-warm, Manual and Skin Servo modes
  • LED Display of Skin Temp., Air Temp., Set Temp. (Resolution 0.1°C) & Heater Power %.
  • LCD to display modes of operation, Alarm messages, Skin Temperature in °F and Timer.
  • Thermistor based sensor with Accuracy of +/- 0.2°C (Temp. Calibration not require)
  • Modes of Operation: Skin Servo, Manual and Pre-Warm
  • Keypad lock facility
  • Skin mode: Temperature range: 34 to 38°C , Resolution : 0.1°C
  • Manual mode: Adjustable Heater Power 0 to 100% (with an increment of 5%), Time duration:15mins
  • Pre-warm mode: To pre-warm the cradle before baby is placed on the bed. (No alarm will be activated during Prewarm mode). Adjustable Heater Power 0 to 25% (with an increment of 5%), Time duration:30mins
  • Overhead heater box: Swivelling, Consists of Ceramic heater (650 Watts) placed in parabolic reflector with a safety grill & Halogen observation light (12V/50W) for observing the baby
  • Operating Voltage: 220-240 VAC, 50Hz, Power: 800W (Max.)
  • Alarm: Audio alarm with 10 minute mute facility & Visual alarms with message on LCD as follows
Alarm Condition Alarm Messages on LCD
Skin Sensor is unplugged /Faulty Skin Sensor Fail
Measured skin temperature higher than set temperature by 1.0°C Skin Temp. High
Measured skin temperature lower than set temperature by 1.0°C Skin Temp. Low
Measured skin temperature higher than 38.0°C (With cut-off) Skin Over Range
Measured skin temperature lower than 34.0°C Skin Under Range
Heater power is 100% for more than 15mins Chk Sens Position
Heater is unplugged or faulty Heater Open
Heater Power drive fail/uncontrollable Chk. Pwr. Drive
Power Fail Alarm: Audio with LED indicator

NIC722 - Infant Radiant Warmer with Fixed Baby Cradle & Drawers with Microprocessor based temperature controller with 4 mode Skin / Air / Manual / Prewarm
Technical Specification:
Heater Box
  • Consists of Ceramic Infra Red Heater 650Watts,
  • The source box can be swivelled to take X-ray
  • Due to temp. insulated cover on the heater box, its exterior surface remains cool even after long run
  • The construction of the stainless steel reflector and Guard rail is such that the complete cleaning of heater & reflector is possible without any electrical shock hazards
Halogen Examination Lamp (12 Volts, 50 Watts)
  • To Provide accurate assessment of Infant colour and sufficient illumination during procedures
Baby Tray (Stainless Steel)
  • Clear acrylic collapsible side panels
  • Stainless steel baby tray with transparent slot to incorporate the facility of under surface phototherapy
  • Acrylic tray and cushion mattress with head up/down facility
  • Acrylic X-ray tray with X-ray guide for positioning the cassette
  • The design aspects also allow ease of dismantling baby care area during cleaning or fumigation
Essential Attachments
  • Lower Shelf and 3 Drawer Storage Cabinet
  • Side Mounted Monitor tray, Pole for Syringe Pump / Infusion Pump
  • S.S. IV Stand.
Micro Computer Based Temp. Controllers
  • To Prove Highest Degree of Accuracy of maintenance of baby’s body Temperature.
Sensors
  • Thermistor type sensor with Accuracy: ± 0.2°C
Alarms
  • Audio alarm mute facility.
  • Mute timing adjustable
Displays
  • 1” Bright 7 segment LED’s (3 nos.) for Skin temp.
  • 0.5” Bright 7 segment LED’s (3 nos.) for Set temp.
  • 0.5” Bright 7 segment LED’s (3 nos.) for Heater Power%.
  • 0.5” Bright 7 segment LED’s (3 nos.) for Air temp.
  • Resolution: 0.1°C
Status: On LCD
  • Control Mode
  • Constant Skin temp in ° Fahrenheit.
  • Timer: Down Timer / APGAR Timer
  • Alarm Messages for
    • Skin Over temperature: Temp > 38.0°C
    • Skin Under temperature: Temp < 34.0°C
    • Skin Positive band: Temp > Set Temp by 1.0°C
    • Skin Negative Band: Temp < Set Temp by 1.0°C
    • Air Over temperature: Temp > 39.0°C
    • Air Under temperature: Temp < 20.0°C
    • Air Positive band: Temp > Set Temp by 1.5°C
    • Air Negative Band: Temp > Set Temp by 3.0°C
    • Heater Failure: Heater connection is failed/damage/faulty
    • System Failure: Fluctuation of voltage supply.
    • Check Sensor position: No proper voltage supply

NIC723 - Infant Radiant Warmer with Fixed Baby Cradle & Drawers with Microprocessor based temperature controller with 3 mode Skin / Manual / Prewarm
Technical Specification:
Heater Box
  • Consists of Ceramic Infra Red Heater 650Watts,
  • The source box can be swivelled to take X-ray
  • Due to temp. insulated cover on the heater box, its exterior surface remains cool even after long run
  • The construction of the stainless steel reflector and Guard rail is such that the complete cleaning of heater & reflector is possible without any electrical shock hazards
Halogen Examination Lamp (12 Volts, 50 Watts)
  • To Provide accurate assessment of Infant colour and sufficient illumination during procedures
Baby Tray (Stainless Steel)
  • Clear acrylic collapsible side panels
  • Stainless steel baby tray with transparent slot to incorporate the facility of under surface phototherapy
  • Acrylic tray and cushion mattress with head up/down facility
  • Acrylic X-ray tray with X-ray guide for positioning the cassette
  • The design aspects also allow ease of dismantling baby care area during cleaning or fumigation
Essential Attachments
  • Lower Shelf and 3 Drawer Storage Cabinet
  • Side Mounted Monitor tray, Pole for Syringe Pump / Infusion Pump
  • S.S. IV Stand.
Micro Computer Based Temp. Controllers
  • To Prove Highest Degree of Accuracy of maintenance of baby’s body Temperature.
Sensors
  • Thermistor type sensor with Accuracy: ± 0.2°C
Alarms
  • Audio alarm mute facility.
  • Mute timing adjustable
Range
  • Skin Mode: 30.0°C to 38.0°C
  • Manual Mode: Vary Heater Power % from 0% to 100% with increment of5%
  • Time duration (Manual mode): 15 mins (factory default)
  • Prewarm mode: Vary Heater Power % from 0% to 25% with increment of5%
  • Time duration (Manual mode): 30 mins (factory default)
Displays
  • 1” Bright 7 segment LED’s (3 nos.) for Skin temp.
  • 0.5” Bright 7 segment LED’s (3 nos.) for Set temp.
  • 0.5” Bright 7 segment LED’s (3 nos.) for Heater Power%.
  • Resolution: 0.1°C
Status: On LCD
  • Control Mode
  • Constant Skin temp in ° Fahrenheit.
  • Timer: Down Timer / APGAR Timer
  • Alarm Messages for
    • Skin Over temperature: Temp > 38.0°C
    • Skin Under temperature: Temp < 34.0°C
    • Skin Positive band: Temp > Set Temp by 1.0°C
    • Skin Negative Band: Temp < Set Temp by 1.0°C
    • Heater Failure: Heater connection is failed/damage/faulty
    • System Failure: Fluctuation of voltage supply.
    • Check Sensor position: No proper voltage supply

NIC732 - Infant Radiant Warmer with Fixed Baby Cradle with Microprocessor based temperature controller with 4 mode Skin / Air / Manual / Prewarm
Technical Specification:
Heater Box
  • Consists of Ceramic Infra Red Heater 650Watts,
  • The source box can be swivelled to take X-ray
  • Due to temp. insulated cover on the heater box, its exterior surface remains cool even after long run
  • The construction of the stainless steel reflector and Guard rail is such that the complete cleaning of heater & reflector is possible without any electrical shock hazards
Halogen Examination Lamp (12 Volts, 50 Watts)
  • To Provide accurate assessment of Infant colour and sufficient illumination during procedures
Baby Tray (Stainless Steel)
  • Clear acrylic collapsible side panels
  • Stainless steel baby tray with transparent slot to incorporate the facility of under surface phototherapy
  • Acrylic tray and cushion mattress with head up/down facility
  • Acrylic X-ray tray with X-ray guide for positioning the cassette
  • The design aspects also allow ease of dismantling baby care area during cleaning or fumigation
Essential Attachments
  • Lower Shelves
  • Side Mounted Monitor tray, Pole for Syringe Pump / Infusion Pump
  • S.S. IV Stand.
Micro Computer Based Temp. Controllers
  • To Prove Highest Degree of Accuracy of maintenance of baby’s body Temperature.
Sensors
  • Thermistor type sensor with Accuracy: ± 0.2°C
Alarms
  • Audio alarm mute facility.
  • Mute timing adjustable
Range
  • Skin Mode: 30.0°C to 38.0°C
  • Air Mode: Room temp. 20.0°C to 39.0°C
  • Manual Mode: Vary Heater Power % from 0% to 100% with increment of 5%
  • Time duration (Manual mode): 15 mins (factory default)
  • Prewarm mode: Vary Heater Power % from 0% to 25% with increment of 5%
  • Time duration (Manual mode): 30 mins (factory default)
Displays
  • 1” Bright 7 segment LED’s (3 nos.) for Skin temp.
  • 0.5” Bright 7 segment LED’s (3 nos.) for Set temp.
  • 0.5” Bright 7 segment LED’s (3 nos.) for Heater Power%.
  • 0.5” Bright 7 segment LED’s (3 nos.) for Air temp.
  • Resolution: 0.1°C
Status: On LCD
  • Control Mode
  • Constant Skin temp in ° Fahrenheit.
  • Timer: Down Timer / APGAR Timer
  • Alarm Messages for
    • Skin Over temperature: Temp > 38.0°C
    • Skin Under temperature: Temp < 34.0°C
    • Skin Positive band: Temp > Set Temp by 1.0°C
    • Skin Negative Band: Temp < Set Temp by 1.0°C
    • Air Over temperature: Temp > 39.0°C
    • Air Under temperature: Temp < 20.0°C
    • Air Positive band: Temp > Set Temp by 1.5°C
    • Air Negative Band: Temp > Set Temp by 3.0°C
    • Heater Failure: Heater connection is failed/damage/faulty
    • System Failure: Fluctuation of voltage supply.
    • Check Sensor position: No proper voltage supply

NIC733 - Infant Radiant Warmer with Fixed Baby Cradle with Microprocessor based temperature controller with 3 mode Skin / Manual / Prewarm
Technical Specification:
Heater Box
  • Consists of Ceramic Infra Red Heater 650Watts,
  • The source box can be swivelled to take X-ray
  • Due to temp. insulated cover on the heater box, its exterior surface remains cool even after long run
  • The construction of the stainless steel reflector and Guard rail is such that the complete cleaning of heater & reflector is possible without any electrical shock hazards
Halogen Examination Lamp (12 Volts, 50 Watts)
  • To Provide accurate assessment of Infant colour and sufficient illumination during procedures
Baby Tray (Stainless Steel)
  • Clear acrylic collapsible side panels
  • Stainless steel baby tray with transparent slot to incorporate the facility of under surface phototherapy
  • Acrylic tray and cushion mattress with head up/down facility
  • Acrylic X-ray tray with X-ray guide for positioning the cassette
  • The design aspects also allow ease of dismantling baby care area during cleaning or fumigation
Essential Attachments
  • Lower Shelves
  • Side Mounted Monitor tray, Pole for Syringe Pump / Infusion Pump
  • S.S. IV Stand.
Micro Computer Based Temp. Controllers
  • To Prove Highest Degree of Accuracy of maintenance of baby’s body Temperature.
Sensors
  • Thermistor type sensor with Accuracy: ± 0.2°C
Alarms
  • Audio alarm mute facility.
  • Mute timing adjustable
Range
  • Skin Mode: 30.0°C to 38.0°C
  • Manual Mode: Vary Heater Power % from 0% to 100% with increment of 5%
  • Time duration (Manual mode): 15 mins (factory default)
  • Prewarm mode: Vary Heater Power % from 0% to 25% with increment of 5%
  • Time duration (Manual mode): 30 mins (factory default)
Displays
  • 1” Bright 7 segment LED’s (3 nos.) for Skin temp.
  • 0.5” Bright 7 segment LED’s (3 nos.) for Set temp.
  • 0.5” Bright 7 segment LED’s (3 nos.) for Heater Power%.
  • Resolution: 0.1°C
Status: On LCD
  • Control Mode
  • Constant Skin temp in ° Fahrenheit.
  • Timer: Down Timer / APGAR Timer
  • Alarm Messages for
    • Skin Over temperature: Temp > 38.0°C
    • Skin Under temperature: Temp < 34.0°C
    • Skin Positive band: Temp > Set Temp by 1.0°C
    • Skin Negative Band: Temp < Set Temp by 1.0°C
    • Heater Failure: Heater connection is failed/damage/faulty
    • System Failure: Fluctuation of voltage supply.
    • Check Sensor position: No proper voltage supply

NIC652 - Infant Radiant Warmer with Detachable Baby Cradle with Microprocessor based temperature controller with 4 mode Skin / Air / Manual / Prewarm
Technical Specification:
Heater Box
  • Consists of Ceramic Infra Red Heater 650Watts,
  • The source box can be swivelled to take X-ray
  • Due to temp. insulated cover on the heater box, its exterior surface remains cool even after long run
  • The construction of the stainless steel reflector and Guard rail is such that the complete cleaning of heater & reflector is possible without any electrical shock hazards
Halogen Examination Lamp (12 Volts, 50 Watts)
  • To Provide accurate assessment of Infant colour and sufficient illumination during procedures
Baby Tray (Stainless Steel)
  • Clear acrylic collapsible side panels
  • Stainless steel baby tray with transparent slot to incorporate the facility of under surface phototherapy
  • Acrylic tray and cushion mattress with head up/down facility
  • Acrylic X-ray tray with X-ray guide for positioning the cassette
  • The design aspects also allow ease of dismantling baby care area during cleaning or fumigation
Essential Attachments
  • Lower Shelves
  • Side Mounted Monitor tray, Pole for Syringe Pump / Infusion Pump
  • S.S. IV Stand.
Micro Computer Based Temp. Controllers
  • To Prove Highest Degree of Accuracy of maintenance of baby’s body Temperature.
Sensors
  • Thermistor type sensor with Accuracy: ± 0.2°C
Alarms
  • Audio alarm mute facility.
  • Mute timing adjustable
Range
  • Skin Mode: 30.0°C to 38.0°C
  • Air Mode: Room temp. 20.0°C to 39.0°C
  • Manual Mode: Vary Heater Power % from 0% to 100% with increment of 5%
  • Time duration (Manual mode): 15 mins (factory default)
  • Prewarm mode: Vary Heater Power % from 0% to 25% with increment of 5%
  • Time duration (Manual mode): 30 mins (factory default)
Displays
  • 1” Bright 7 segment LED’s (3 nos.) for Skin temp.
  • 0.5” Bright 7 segment LED’s (3 nos.) for Set temp.
  • 0.5” Bright 7 segment LED’s (3 nos.) for Heater Power%.
  • 0.5” Bright 7 segment LED’s (3 nos.) for Air temp.
  • Resolution: 0.1°C
Status: On LCD
  • Control Mode
  • Constant Skin temp in ° Fahrenheit.
  • Timer: Down Timer / APGAR Timer
  • Alarm Messages for
    • Skin Over temperature: Temp > 38.0°C
    • Skin Under temperature: Temp < 34.0°C
    • Skin Positive band: Temp > Set Temp by 1.0°C
    • Skin Negative Band: Temp < Set Temp by 1.0°C
    • Air Over temperature: Temp > 39.0°C
    • Air Under temperature: Temp < 20.0°C
    • Air Positive band: Temp > Set Temp by 1.5°C
    • Air Negative Band: Temp > Set Temp by 3.0°C
    • Heater Failure: Heater connection is failed/damage/faulty
    • System Failure: Fluctuation of voltage supply.

NIC653 - Infant Radiant Warmer with Detachable Baby Cradle with Microprocessor based temperature controller with 3 mode Skin / Manual / Prewarm
Technical Specification:
Heater Box
  • Consists of Ceramic Infra Red Heater 650Watts,
  • The source box can be swivelled to take X-ray
  • Due to temp. insulated cover on the heater box, its exterior surface remains cool even after long run
  • The construction of the stainless steel reflector and Guard rail is such that the complete cleaning of heater & reflector is possible without any electrical shock hazards
Halogen Examination Lamp (12 Volts, 50 Watts)
  • To Provide accurate assessment of Infant colour and sufficient illumination during procedures
Baby Tray (Stainless Steel)
  • Clear acrylic collapsible side panels
  • Stainless steel baby tray with transparent slot to incorporate the facility of under surface phototherapy
  • Acrylic tray and cushion mattress with head up/down facility
  • Acrylic X-ray tray with X-ray guide for positioning the cassette
  • The design aspects also allow ease of dismantling baby care area during cleaning or fumigation
Essential Attachments
  • Lower Shelves
  • Side Mounted Monitor tray, Pole for Syringe Pump / Infusion Pump
  • S.S. IV Stand.
Micro Computer Based Temp. Controllers
  • To Prove Highest Degree of Accuracy of maintenance of baby’s body Temperature.
Sensors
  • Thermistor type sensor with Accuracy: ± 0.2°C
Alarms
  • Audio alarm mute facility.
  • Mute timing adjustable
Range
  • Skin Mode: 30.0°C to 38.0°C
  • Manual Mode: Vary Heater Power % from 0% to 100% with increment of 5%
  • Time duration (Manual mode): 15 mins (factory default)
  • Prewarm mode: Vary Heater Power % from 0% to 25% with increment of 5%
  • Time duration (Manual mode): 30 mins (factory default)
Displays
  • 1” Bright 7 segment LED’s (3 nos.) for Skin temp.
  • 0.5” Bright 7 segment LED’s (3 nos.) for Set temp.
  • 0.5” Bright 7 segment LED’s (3 nos.) for Heater Power%.
  • Resolution: 0.1°C
Status: On LCD
  • Control Mode
  • Constant Skin temp in ° Fahrenheit.
  • Timer: Down Timer / APGAR Timer
  • Alarm Messages for
    • Skin Over temperature: Temp > 38.0°C
    • Skin Under temperature: Temp < 34.0°C
    • Skin Positive band: Temp > Set Temp by 1.0°C
    • Skin Negative Band: Temp < Set Temp by 1.0°C
    • Heater Failure: Heater connection is failed/damage/faulty
    • System Failure: Fluctuation of voltage supply.
    • Check Sensor position: No proper voltage supply

NIC642 - Infant Radiant Warmer with Detachable Baby Cradle with Microprocessor based temperature controller with 4 mode Skin / Air / Manual / Prewarm
Technical Specification:
Heater Box
  • Consists of Ceramic Infra Red Heater 650Watts,
  • The source box can be swivelled to take X-ray
  • Due to temp. insulated cover on the heater box, its exterior surface remains cool even after long run
  • The construction of the stainless steel reflector and Guard rail is such that the complete cleaning of heater & reflector is possible without any electrical shock hazards
Halogen Examination Lamp (12 Volts, 50 Watts)
  • To Provide accurate assessment of Infant colour and sufficient illumination during procedures
Baby Tray (Stainless Steel)
  • Clear acrylic collapsible side panels
  • Stainless steel baby tray with transparent slot to incorporate the facility of under surface phototherapy
  • Acrylic tray and cushion mattress with head up/down facility
  • Acrylic X-ray tray with X-ray guide for positioning the cassette
  • The design aspects also allow ease of dismantling baby care area during cleaning or fumigation
Essential Attachments
  • Side Mounted Monitor tray, Pole for Syringe Pump / Infusion Pump
  • S.S. IV Stand.
Micro Computer Based Temp. Controllers
  • To Prove Highest Degree of Accuracy of maintenance of baby’s body Temperature.
Sensors
  • Thermistor type sensor with Accuracy: ± 0.2°C
Alarms
  • Audio alarm mute facility.
  • Mute timing adjustable
Range
  • Skin Mode: 30.0°C to 38.0°C
  • Air Mode: Room temp. 20.0°C to 39.0°C
  • Manual Mode: Vary Heater Power % from 0% to 100% with increment of 5%
  • Time duration (Manual mode): 15 mins (factory default)
  • Prewarm mode: Vary Heater Power % from 0% to 25% with increment of 5%
  • Time duration (Manual mode): 30 mins (factory default)
Displays
  • 1” Bright 7 segment LED’s (3 nos.) for Skin temp.
  • 0.5” Bright 7 segment LED’s (3 nos.) for Set temp.
  • 0.5” Bright 7 segment LED’s (3 nos.) for Heater Power%.
  • 0.5” Bright 7 segment LED’s (3 nos.) for Air temp.
  • Resolution: 0.1°C
Status: On LCD
  • Control Mode
  • Constant Skin temp in ° Fahrenheit.
  • Timer: Down Timer / APGAR Timer
  • Alarm Messages for
    • Skin Over temperature: Temp > 38.0°C
    • Skin Under temperature: Temp < 34.0°C
    • Skin Positive band: Temp > Set Temp by 1.0°C
    • Skin Negative Band: Temp < Set Temp by 1.0°C
    • Air Over temperature: Temp > 39.0°C
    • Air Under temperature: Temp < 20.0°C
    • Air Positive band: Temp > Set Temp by 1.5°C
    • Air Negative Band: Temp > Set Temp by 3.0°C
    • Heater Failure: Heater connection is failed/damage/faulty
    • System Failure: Fluctuation of voltage supply.
    • Check Sensor position: No proper voltage supply

NIC643 - Infant Radiant Warmer with Detachable Baby Cradle with Microprocessor based temperature controller with 3 mode Skin / Manual / Prewarm
Technical Specification:
Heater Box
  • Consists of Ceramic Infra Red Heater 650Watts,
  • The source box can be swivelled to take X-ray
  • Due to temp. insulated cover on the heater box, its exterior surface remains cool even after long run
  • The construction of the stainless steel reflector and Guard rail is such that the complete cleaning of heater & reflector is possible without any electrical shock hazards
Halogen Examination Lamp (12 Volts, 50 Watts)
  • To Provide accurate assessment of Infant colour and sufficient illumination during procedures
Baby Tray (Stainless Steel)
  • Clear acrylic collapsible side panels
  • Stainless steel baby tray with transparent slot to incorporate the facility of under surface phototherapy
  • Acrylic tray and cushion mattress with head up/down facility
  • Acrylic X-ray tray with X-ray guide for positioning the cassette
  • The design aspects also allow ease of dismantling baby care area during cleaning or fumigation
Essential Attachments
  • Side Mounted Monitor tray, Pole for Syringe Pump / Infusion Pump
  • S.S. IV Stand.
Micro Computer Based Temp. Controllers
  • To Prove Highest Degree of Accuracy of maintenance of baby’s body Temperature.
Sensors
  • Thermistor type sensor with Accuracy: ± 0.2°C
Alarms
  • Audio alarm mute facility.
  • Mute timing adjustable
Range
  • Skin Mode: 30.0°C to 38.0°C
  • Manual Mode: Vary Heater Power % from 0% to 100% with increment of 5%
  • Time duration (Manual mode): 15 mins (factory default)
  • Prewarm mode: Vary Heater Power % from 0% to 25% with increment of 5%
  • Time duration (Manual mode): 30 mins (factory default)
Displays
  • 1” Bright 7 segment LED’s (3 nos.) for Skin temp.
  • 0.5” Bright 7 segment LED’s (3 nos.) for Set temp.
  • 0.5” Bright 7 segment LED’s (3 nos.) for Heater Power%.
  • Resolution: 0.1°C
Status: On LCD
  • Control Mode
  • Constant Skin temp in ° Fahrenheit.
  • Timer: Down Timer / APGAR Timer
  • Alarm Messages for
    • Skin Over temperature: Temp > 38.0°C
    • Skin Under temperature: Temp < 34.0°C
    • Skin Positive band: Temp > Set Temp by 1.0°C
    • Skin Negative Band: Temp < Set Temp by 1.0°C
    • Heater Failure: Heater connection is failed/damage/faulty
    • System Failure: Fluctuation of voltage supply.
    • Check Sensor position: No proper voltage supply